Opentrons Flex 96-Channel Pipette


The Opentrons Flex 96-Channel pipette enables accurate high throughput pipetting, optimized for working in 96-well plates. Flex Pipettes use air displacement technology to offer highly accurate pipetting with pipette volume ranges from 1 – 1000 μL.


Smart sensors support automatic calibration, real-time positioning, and error detection. The 96-channel pipette occupies both pipette mounts and can easily be removed and swapped to an alternate pipette configurations to adapt your workflow needs.


Four 96-Channel Tip Rack Adapters are included with the purchase of one Flex 96-Channel Pipette.

*Now accepting pre-orders for the 1-200 μL Opentrons Flex 96-Channel Pipette, shipping Q3 2025*

Prices listed here are valid in the US and select other territories. Please get in touch with your local team for pricing for your territory.

999-00196-0
Volume Range

1-Year Warranty

Ships in 7 business days

Create a quote after you add to cart
Create a quote after you add to cart

Features

Accurate pipetting across 96 channels

The Flex 96-Channel Pipette uses air displacement technology to offer highly accurate pipetting from 1 to 1000 µL

Modular pipetting heads

Install or swap out your pipettes yourself in minutes with minimal effort

Automated calibration

Smart sensors support automatic calibration, real-time positioning, and error detection

Tips built for automation

Automation-optimized tips designed specifically for the Flex

Tech Specs

Accuracy & precision (1-200 μl)
Tip Size Volume %D %CV
50μl 1μl ±10% ±6%
50μl 5μl ±4% ±2%
50μl 50μl ±1.5% ±1%
200μl 200μl ±1% ±1%
Accuracy & precision (5-1,000 μl)
Tip Size Volume %D %CV
50μl 5μl ±10% ±5%
50μl 50μl ±2.5% ±1.25%
200μl 200μl ±1.5% ±1.25%
1,000μl 1,000μl ±1.5% ±1.5%

Resources

Interactive Demos

FAQs

To pick up tips, The Opentrons Flex 96-Channel Pipette requires tip racks be in a 96-Channel Tip Rack Adapter. Four 96-Channel Tip Rack Adapters are included with the purchase of one Flex 96-Channel Pipette.

Yes, Opentrons Flex users will be able to offset the 96-channel pipette head to pick up 8 tips (1 column). This functionality can only be achieved through the Python API.

Yes, Flex pipettes include pressure sensors for detecting liquid height. These sensors are contained within the A1 and H12 channels on the 96-channel pipette, and the outermost channels on the 8-channel pipette (1 & 8).

The Opentrons Flex 8-channel and 96-channel pipettes will aspirate and dispense the same volume with all channels at the same time. Volume is not independently controlled across channels.

OT-2 pipettes and OT-2 pipette tips are not compatible with the Opentrons Flex. The OT-2 pipette nose cone and Flex pipette nose cones are constructed differently, preventing cross-compatibility. OT-2 tip boxes/refills are also not compatible with the Flex.

Opentrons can only guarantee performance of the Flex when using our Flex pipette tips.