The OT-2: Your Personal Pipetting Robot

Get started automating your lab today

ot2 photo

Lab automation has never been easier.

The OT-2 is a bench-top liquid handler designed to be accessible and flexible enough to automate many common applications.

Community

“

Our life is so much easier now because the OT-2 is really good.

– Marilene Pavan, Boston University

BOSTON
“

There are days we have Opentrons transfer 800 samples. I would never be able to do as many samples by hand.

– Kaja Wasik, Gencove

GENCOVE
“

Before, I had to leave one whole day for one ELISA run. Now, the robot frees my hands.

– Annabel Shang, MilliporeSigma

Millpore
“

We are currently working with our finance folks to acquire two additional OT-2 instruments, because we LOVE this platform.

– Adam Ericsen, Emory University

emory

Small Footprint, Big Impact

Fits on half a standard lab bench

Metric (width, depth, height)
63 cm x 57 cm x 66 cm
Imperial (width, depth, height)
25 in x 22.5 in x 26 in

Performs like robots costing 10x more

Accuracy
15% at 1 μL
Precision
5% at 1 μL

Getting Started

Everything you need to automate your lab, including support at every step

Demo

Demo the OT-2

Get a virtual demo of the OT-2 and get answers about your specific application

Cart

Order the OT-2

Bundle your OT-2 with the pipettes, hardware and labware you need

Receive and unbox your OT-2

Easy to unbox and set up yourself using our online guide

Custom Protocols

Custom or urgent support?

Contact us for onsite installation, custom protocol development and more!

hardware animation 6467a7d8ae4ea24cea0766c5302fcc72

Use what you have, add what you need

Choose your hardware and pipettes

11 deck slots, plus trash, and space for two swappable, single and 8-Channel pipettes

Magnetic, Temp, Thermocycler, Heater-Shaker, and HEPA Filter modules available

Use your existing labware and reagents

Reagent and consumables agnostic

Compatible with any ANSI/SLAS-compliant labware

Choose the protocol method that’s right for you

A protocol is the code that runs the robot through your experiment. We’ve built different ways to create a protocol that are accessible for everyone – from the automation rookie to the experienced Python coder.

Custom Protocols

Custom Protocol

Best for users that want a unique or complex protocol – fast

Get our experts to create a protocol optimized for your experiment

Protocol Library V1

Protocol Library

Best for new users looking to gain experience

Browse well-documented, open-source protocols

Protocol Designer

Protocol Designer

Best for users who want to create a new protocol – without having to code

Use our intuitive drag-and-drop interface to create new protocols

Pyton API

Python Protocol API

Best for users with Python experience

Write your own protocol using our open-source Python Protocol API

ot2 pipettes

Space for two pipettes

Reliable liquid transfers from 1–1000 µL with our swappable single- and 8-channel pipettes

ot2 pipettes single

SINGLE-CHANNEL

P20
P300
P1000

ot2 pipettes 8 channel

8-Channel

P20
P300

Support & Services

Need help? We’ve got you covered

CSM 1

On-call support offers real-time chat support on your desktop or mobile

Knowledge Hub

Visit the Knowledge Hub to Browse our in-depth support articles and videos

Custom Protocols

Have our experts develop a Custom Protocol for your application

Services 1

Warranties and services maximize your lab’s performance

Tech Specs

ot2 tech specs
Dimensions

63 cm x 57 cm x 66 cm / 25 in x 22.5 in x 26 in (W,D,H)
Takes up half of your standard lab bench

SPECSOT2

Weight

48 kg or 105.8 lb

Pipette configurations

Single- and 8-channel pipetting
2 pipette mounts, for a configuration of 1 or 2 single- or 8-channel pipettes. Pipettes are easily interchangeable.

Pipette volumes and specs
Deck

11 deck slots enable countless configurations
Deck slots are compatible with standard SBS dimensions.
Deck also includes a removable trash bin.

Connectivity

WiFi 2.4 GHz IEEE 802.11b/g/n, USB 2.0

Video

Integrated video camera (Included for future or custom functionality)

Audio

Integrated audio system (Included for future or custom functionality)

Power requirements

Robot Power Input: 36 VDC, 6.1 A
Power Adapter Input: 100-240 VAC, 50/60 Hz, 4.0 A/115 VAC, 2.0 A/230 VAC
For additional power requirement information, please review this support article

Frame composition

Rigid steel and CNC aluminum design

Operating temperature & relative humidity

Recommended 20-24 °C
Recommended 40-60% RH

Time to fill 96-well plate
Pipette Distribute,
one tip
Transfer,
same tip
Transfer,
new tip
P20 8-Channel 0:22 1:35 4:19
P300 8-Channel 0:26 1:38 4:24
P20 Single-Channel 3:12 13:10 36:01
P300 Single-Channel 4:01 13:39 41:31
P1000 5:54 15:21 44:50
Single-channel accuracy & precision
Pipette Volume Random Error Systematic Error
%D µl %CV µl
P20 1
10
20
±15%
±2%
±1.5%
0.15µl
0.2µl
0.3µl
±5%
±1%
±0.8%
0.05µl
0.1µl
0.16µl
P300 20
150
300
±4%
±1%
±0.6%
0.8µl
1.5µl
1.8µl
±2.5%
±0.4%
±0.3%
0.05µl
0.6µl
0.9µl
P1000 100
500
1000
±2%
±1%
±0.7%
2.0µl
5.0µl
7.0µl
±1%
±0.2%
±0.15%
1µl
1µl
1.5µl
8-channel accuracy & precision
Pipette Volume Random Error Systematic Error
%D µl %CV µl
P20 1
10
20
±20%
±3%
±2.2%
0.2µl
0.3µl
0.44µl
±10%
±2%
±1.5%
0.1µl
0.2µl
0.3µl
P300 20
150
300
±10%
±2.5%
±1.5%
2.0µl
3.75µl
4.5µl
±4%
±0.8%
±0.5%
0.8µl
1.2µl
1.5µl
Sterility

Full polycarbonate enclosure designed to limit exposure. The OT-2 is not a sterile environment
Add the Opentrons HEPA Module, or add your own UV lights

Viscosity
For best practices on automating viscous liquid handling, review our application note
API compatibility

Interfaces with any product that has an API
Opentrons software is open-source and users can make adjustments to the software as needed to interface with other products. While all integrations are possible, they all require custom code work i.e. most customers will need to write custom libraries and install software packages that are not shipped with the robot.

Interface
Windows 10, macOS 10.10 or later, Ubuntu 12.04 or later.
Tip compatibility

Use Opentrons Tips and Opentrons Filter Tips
Opentrons guarantees the performance of Opentrons Tips which were specifically designed for the OT-2. The OT-2 was designed to accommodate all universal bore tip types, but tip quality and dimensions can vary significantly so third party tip performance cannot be guaranteed by Opentrons.

Labware compatibility

Use common labware already in our Labware Library, or add your own. If your labware fits on the OT-2 but is not in the Labware Library, it can be added by you using our Labware Creator, or by our team.

Warranty

One year warranty against manufacturer defects
Stress tested to last a minimum of 3 years of full-time use

Return policy
60 day return policy
Certifications

Certifications complete (CE, FCC, NRTL, CB, ISO 9001)
Not certified/validated (IVD, GMP)

Example Commands

Tip Handling

  • Pick up tip
  • Drop tip
  • Return tip
  • New tip (always, never, once)
  • More…

Liquid Control
  • Aspirate
  • Dispense
  • Aspirate speed
  • Dispense speed
  • Blowout
  • Touch tip
  • Mix
  • Air gap

Utility

  • Move to
  • Delay
  • User-specified pause
  • Homing
  • Comment


Transfer

  • Basic transfer
  • Volumes > Pipette Volume
  • Distribute
  • Consolidate


Variable Settings

  • Speed
  • Volume
  • Count


Other

  • Dispose
What’s in the box

BOX

OT-2 robot
Calibration block
Removable block
Trash container
Quickstart guide
Power supply and regional IEC cable
Ethernet cable
Ethernet-to-USB dongle
Tool kit

Introduction to the OT-2

Frequently Asked Questions

To view FAQs for all products, click here.

Yes, the Opentrons Flex’s side windows and front door are removable.

The Heater-Shaker Module can assist in automating a variety of wetlab workflows. Here is an example of a DNA Extraction protocol leveraging the Heater-Shaker Module.

Yes, this can be achieved through the Opentrons app. You will need to be connected to a computer that is on the same WiFi or wired network as the liquid handler.

Yes. Just like the OT-2, the Flex has full capability to accept custom or irregular labware. Our custom labware creation tools can be used to help create the compatible labware definitions that are required to use irregular labware with our software.

It is possible to use multiple modules of the same type on the OT-2. This is true for our Temperature Module, Magnetic Module, and Heater-Shaker Module. For a walkthrough on setting this up for your OT-2 and considerations when using multiple modules, please reference this guide.

You can use one computer running the Opentrons App to connect to multiple robots. This is true for both the Opentrons Flex and OT-2 robots.

OT-2 pipettes and OT-2 pipette tips are not compatible with the Opentrons Flex. The OT-2 pipette nose cone and Flex pipette nose cones are constructed differently, preventing cross-compatibility. OT-2 tip boxes/refills are also not compatible with the Flex.

Opentrons can only guarantee performance of the Flex when using our Flex pipette tips.

No, Opentrons’ OT-2 pipettes will not be compatible with the Opentrons Flex. The Opentrons Flex is intended for use with our Flex Pipettes.

Yes, Flex pipettes include pressure sensors for detecting liquid height. These sensors are contained within the A1 and H12 channels on the 96-channel pipette, and the outermost channels on the 8-channel pipette (1 & 8).

Labs within the Opentrons community have installed the OT-2 pipetting robot in biological safety cabinets (as well as in inert atmosphere glove boxes), including Labconco’s Logic Vue Class II enclosures and Baker’s AeroPROTECT 360 enclosures. We recommend you cross-check the dimensions of the OT-2 with those of your chosen biological safety cabinet.

Users do not need a computer connected to the Opentrons Flex to run a protocol. Users will need to build the protocol on a computer and upload it to the Flex. Once uploaded, the protocol can be ran solely off the display.

You do not need to know how to code to use the Opentrons Flex platform. Protocols can be created using Opentrons’ plug-and-play protocol design tools. We also have an Applications Engineering team that can build custom protocols for you.

You do not need to know how to code to use the Opentrons platform. Protocols can be created using Opentrons’ plug-and-play protocol design tools. We also have an Applications Engineering team that can build custom protocols for you.

The Opentrons Flex 8-channel and 96-channel pipettes will aspirate and dispense the same volume with all channels at the same time. Volume is not independently controlled across channels.

Yes, Opentrons Flex users will be able to offset the 96-channel pipette head to pick up 8 tips (1 column). This functionality can only be achieved through the Python API.

Crashes will be detected by the Opentrons Flex by comparing the actual vs. intended move of the embedded motors.

For example, If the robot expected 2 motor rotations, but only 1 rotation is completed due to a crash(which stops the motors from moving), then the robot would detect a discrepancy between intended move and actual. This would then throw an error and let you know you need to recalibrate.

Due to the open-source nature of the robot, many of our users have found success integrating the OT-2 with third party lab equipment.

Integrations such as these would be a DIY project taken on by your lab, and would not fall under the scope of the Opentrons Support team.

Yes, we are still using stepper motors with linear encoders. This enables us to track the steps of the motor and handle errors with positioning.

For those familiar with the homing patterns of the OT-2 robot, we no longer rely on rehoming the axis with limit switches in an open-looped feedback architecture. We can use a closed-loop system, powered by our new encoders to reduce the number of times that we have to home in a protocol.

This is not an optical process. The robot uses capacitance sensors and a calibration probe (shipped with Flex pipettes) to perform automated calibration. Capacitive sensing for automated calibration is done by having a current run through the pipette’s metal tip at the bottom and then through a calibration tip until it’s grounded.

Users can save up to 8 protocols to the Opentrons Flex display.

To pick up tips, The Opentrons Flex 96-Channel Pipette requires tip racks be in a 96-Channel Tip Rack Adapter. Four 96-Channel Tip Rack Adapters are included with the purchase of one Flex 96-Channel Pipette.

Yes! When carrying out multiple protocols on a Heater-Shaker Module, it may be required to have the module heat, but not shake. This option is possible on the Heater-Shaker Module.

Yes. Compatible labware for the Opentrons Flex Magnetic Block includes:

–Opentrons Tough 0.2 mL PCR well plate (full skirt)

–NEST 2 mL 96 Deep Well Plate

The Opentrons Flex requires on-site installation. We also offer both remote and on-site services solutions, as well as annual preventative maintenance.

You can find more information on Opentrons’ service offerings here.

The OT-2 has the following certifications: CE, FCC, NRTL, ISO 9001, CB
Note that the robot is not IVD or GMP certified/validated

You can find our community’s collection of OT-2 protocols within our Protocol Library. Feel free to search the library for a protocol that matches your lab’s workflows!

If you don’t see a protocol that matches your lab’s processes, we offer a couple different options for creating your protocol. Additionally, you can leverage our team of engineers for a custom protocol development.

The Opentrons Flex Magnetic Block GEN1 is compatible with the Opentrons Flex robot, while the OT-2 Magnetic Module GEN2 is compatible with the OT-2 robot.

The Magnetic Block is a plate-sized magnet that relies on the Opentrons Flex Gripper GEN1 to move your plates on and off the block. Conversely, the Magnetic Module is a powered unit that turns the magnets on and off, and is intended for use with a liquid handling robot without a gripper.

20 cm (8 in) of side and back clearance for cables, USB connections, and to dissipate exhaust from instruments like the Thermocycler or Temperature Module.

Out of the box, the OT-2 can support any automation-compatible, ANSI/SLAS compliant well plates, plus a wide range of snap-cap and screw-cap tubes up to a 50 mL conical with our 4-in-1 Tube Rack. You can see the full range of supported labware in our Labware Library.